. 2017 Aug 23;32(3):e22309. doi: 10.1002/jcla.22309

Table 3.

Predonation screening using the noninvasive and invasive POCT methods according to the venous reference value

Donors Noninvasive POCT method Invasive POCT method
Total
Screened donors 339 350
Rejected by venous method 61 66
Rejected 90 84
Rejected correctly 33 52
Rejected falsely too low 57 32
Accepted correctly 192 234
Accepted falsely too high 30 14
Accordance (%) 66.37 81.71
Women
Screened donors 193 204
Rejected by venous method 55 60
Rejected 74 66
Rejected correctly 27 48
Rejected falsely too low 46 18
Accepted correctly 92 126
Accepted falsely too high 28 12
Accordance (%) 61.65 85.29
Men
Screened donors 146 146
Rejected by venous method 6 6
Rejected 16 18
Rejected correctly 6 4
Rejected falsely too low 11 14
Accepted correctly 100 108
Accepted falsely too high 2 2
Accordance (%) 72.60 76.71