. 2017 Apr 11;32(2):e22229. doi: 10.1002/jcla.22229

Table 1.

Icterus interference cut‐off from Roche7 observed values and TCL (see abbreviation in text). For each analyte (column A), the tested range (column C) with the corresponding units (column B) is done. Roche's icterus interference cut‐off at 10% (column D) is compared to the value observed when there is 10% variation (column E). At least, I ict observed at TCL level (column F) is done. The values differ when TCL is not close to 10% (column G). Therefore, the interpretation is adapted to the analytical performance goals of the laboratory

Column A Column B Column C Column D Column E Column F Column G
Tests Units Range tested (min‐max) I ict 10% proposal from Roche I ict observed at 10% variation TCL (%) I ict observed at TCL (column F)
ALT U/L 12‐359 60 >60 11.3 >60
ALB g/L 22.8‐37.9 60 >60 5.3 45
ALP U/L 48‐381 60 >60 7.1 58
AMY U/L 30‐158 60 >60 6.3 >60
AST U/L 14‐734 60 >60 9 >60
CHOLT mmol/L 2.06‐6.72 14 14 4.7 7
CREA μmol/L 55‐912 15 13 6.0 6
FRUC mmol/L 139‐228 5 5 9.5 5
GGT U/L 26‐179 20 >60 8.9 >60
HDLc mmol/L 0.5‐1.74 30 20 6.9 12
Iron μmol/L 5.14‐27 60 >60 14.3 >60
LIP U/L 21‐410 50 >60 13.1 >60
Phos mmol/L 0.73‐1.95 60 >60 5.2 34
PROT g/L 45.2‐67.5 20 16 5.2 6
TG mmol/L 0.81‐3.64 10 20 11 21
UA μmol/L 204‐670 40 43 5.3 20