Table 4.

Diagnostic accuracy of tested analytes on Combur‐10M and iChem Velocity dipsticks

Analyte N iChem Velocity Combur‐10M
Weighted kappa (95% CI) Specificity (%) Sensitivity (%) Accuracy (%) Weighted kappa (95% CI) Specificity (%) Sensitivity (%) Accuracy (%)
Glucose 213 0.943 (0.913 to 0.973) 98.4 100 95 0.926 (0.880 to 0.972) 100 99.5 99
Bilirubin 229 0.785 (0.674 to 0.895) 96.2 72 88 0.805 (0.726 to 0.883) 93.2 80 89
Protein 40 0.795 (0.685 to 0.905) 100 95 95 0.898 (0.835 to 0.961) 75 100 88
Leukocyte esterase 40 0.603 (0.405 to 0.800) 58.3 100 88 0.409 (0.206 to 0.612) 50 100 83
Hemoglobin 229 0.010 (−0.118 to 0.138) 65.6 68.2 50 0.044 (−0.078 to 0.166) 62.7 73.6 51