Skip to main content
. 2018 Feb 25;32(5):e22390. doi: 10.1002/jcla.22390

Table 5.

Ascorbic acid interference on tested analytes on iChem Velocity and Combur‐10M dipsticks

Analyte Analyte concentration (mg/dL) iChem Velocity reported categorya Combur‐10M reported categoryb Ascorbic acid concentration iChem Velocity Combur‐10M
Glucose 50
1+a
1+b
0 Normal 1+
20 Normal Normal
50 Normal Normal
100 Normal Normal
200 Normal Normal
500 Normal Normal
100
2+a
2+b
0 1+ 2+
20 1+ 1+
50 1+ 1+
100 1+ Normal
200 1+ Normal
500 1+ Normal
300
3+a
3+b
0 2+ 3+
20 2+ 2+
50 2+ 3+
100 2+ 2+
200 2+ 2+
500 Normal 2+
Hemoglobin 0.03
2+a
2+b
0 2+ 2+
20 1+ Negative
40 1+ Negative
200 Negative Negative
500 Negative Negative
0.06
3+a
3+b
0 2+ 2+
20 1+ 2+
40 1+ 2+
200 Negative 1+
500 Negative Negative
0.10
3+a
3+b
0 2+ 3+
20 2+ 3+
40 1+ 3+
200 Negative 3+
500 Negative Negative
Nitrite 0.10
Negativea
Negativeb
0 Positive Negative
20 Negative Negative
40 Negative Negative
200 Negative Negative
500 Negative Negative
0.20
Positivea
Positiveb
0 Positive Positive
20 Positive Positive
40 Negative Positive
200 Negative Positive
500 Negative Negative
Bilirubin 1.8
1+a
1+b
0 2+ 3+
20 1+ 2+
40 Negative 1+
200 Negative 1+
500 Negative Negative
4.0
2+a
2+b
0 2+ 3+
20 2+ 3+
40 1+ 2+
200 1+ 2+
500 Negative 1+

Highlighted fields present interference points stated by the manufacturers.