Sensitivity, specificity, and positive and negative predictive values of different systems compared with HPLC before and after harmonization

System Before harmonization After harmonization
ECL‐Roche CLIA‐Diasorin EIA‐DIAsource RIA‐DIAsource ECL‐Roche CLIA‐Diasorin EIA‐DIAsource RIA‐DIAsource
Sensitivity 86.4 CI: 80.2‐90.9 99.3 CI: 95.9‐99.9 78.1 CI: 71.1‐83.7 82.7 CI: 74.9‐88.5 90.3 CI: 84.8‐94.1 93.5 CI: 88.1‐96.6 91.4 CI:86.0‐94.9 87.2 CI: 80.0‐92.1
Specificity 87.9 CI: 78.9‐93.5 47.3 CI: 36.9‐57.8 87.9 CI: 78.9‐93.5 81.7 CI: 72.0‐88.6 83.1 CI: 73.4‐89.9 78.4 CI: 68.5‐86.0 69.2 CI: 58.5‐78.2 72.5 CI: 62.0‐81.1
PPV 93.2 CI: 88.0‐96.4 75.8 CI: 69.2‐81.4 92.6 CI: 86.9‐96.0 86.6 CI: 79.1‐91.7 91.4 CI: 86.0‐94.9 87.8 CI: 81.6‐92.2 85.1 CI: 79.0‐89.7 82.2 CI: 74.7‐87.9
NPV 76.9 CI: 67.4‐84.3 97.7 CI: 86.7‐99.8 67.2 CI: 57.9‐75.3 76.7 CI: 67.0‐84.4 81.3 CI: 71.4‐88.4 87.9 CI: 78.5‐93.7 80.7 CI: 69.9‐88.4 79.5 CI: 68.9‐87.2

NPV, negative predictive value; PPV, positive predictive value.

Vitamin D status was defined based on serum 25(OH)D concentrations (nmol/L): desirable >50, undesirable <50.