Table 4.
ECL‐Roche | CLIA‐Diasorin | EIA‐DIAsource | RIA‐DIAsource | |
---|---|---|---|---|
Sensitivity | 92.1 CI: 86.1‐95.8 | 92.4 CI: 85.7‐96.2 | 90.1 CI: 83.7‐94.3 | 91.8 CI: 84.0‐96.1 |
Specificity | 84.6 CI: 76.8‐90.2 | 83.5 CI: 75.7‐89.3 | 77.2 CI: 68.6‐84.1 | 67.2 CI: 58.1‐75.1 |
PPV | 87.2 CI: 80.5‐91.9 | 83.9 CI: 76.3‐89.5 | 82.0 CI: 74.9‐87.5 | 68.7 CI: 59.9‐76.3 |
NPV | 90.5 CI: 83.2‐94.9 | 92.2 CI: 85.3‐96.1 | 87.1 CI: 79.0‐92.5 | 91.3 CI: 83.1‐95.8 |
Vitamin D status based on circulating 25(OH)D concentration (nmol/L) for HPLC: sufficiency >38.55, insufficiency ≤38.55; ECL‐Roche: sufficiency >43.5, insufficiency ≤43.5; CLIA‐Diasorin: Sufficiency >30.0; insufficiency ≤30.0; EIA‐Diasource: sufficiency >51.2; insufficiency ≤51.2; RIA‐Diasource: Sufficiency >51.4; insufficiency ≤51.4.
To develop HPLC‐specific cutoff points, changes of serum iPTH in different concentrations of serum 25(OH)D were used. Using ROC analysis, these cutoffs were then employed to develop cutoff points for other systems.