. 2018 Mar 30;32(6):e22429. doi: 10.1002/jcla.22429

Table 2.

Inaccuracy of the proposed method for lyophilized and liquid serum specimensa (mmol/L, n = 10)

Cations Specimen IC CV (%) Reference value Bias (%)
Na Serum‐I 132.79 0.36 133.0b −0.16
Serum‐II 148.54 0.20 149.0b −0.31
2015 RELA‐A 140.40 0.18 139.59c 0.58
2015 RELA‐B 154.20 0.13 154.20c 0.00
2016 RELA‐A 130.78 0.20 131.18c −0.29
2016 RELA‐B 141.55 0.24 142.08c −0.34
K Serum‐I 3.444 0.30 3.46b −0.46
Serum‐II 5.360 0.15 5.36b 0.02
2015 RELA‐A 5.880 0.19 5.860c 0.34
2015 RELA‐B 3.130 0.95 3.133c −0.09
2016 RELA‐A 4.917 0.32 4.931c −0.28
2016 RELA‐B 6.807 0.30 6.829c −0.32
Mg Serum‐I 0.732 0.39 0.73b 0.27
Serum‐II 0.980 1.00 0.97b 1.03
2015 RELA‐A 1.943 0.97 1.928c 0.80
2015 RELA‐B 0.911 0.51 0.908c 0.28
2016 RELA‐A 0.732 0.44 0.722c 1.33
2016 RELA‐B 1.468 0.59 1.454c 0.99
Ca Serum‐I 2.037 0.83 2.03b 0.34
Serum‐II 3.009 0.96 2.99b 0.64
2015 RELA‐A 3.029 0.42 3.026c 0.11
2015 RELA‐B 1.992 0.80 2.001c −0.46
2016 RELA‐A 3.464 0.32 3.481c −0.48
2016 RELA‐B 2.513 0.45 2.526c −0.52
a

Limits of equivalence of IFCC‐RELA: Na ± 1.25%; K ± 2.00%; Mg ± 3.75%; Ca ± 2.50%.

b

The reference value certified by National Center for Clinical Laboratories using ICP‐MS.

c

Mean of all participants