Table 1.

Characteristics of assays used for quantitative HBsAg measurement

Device Trade Name (abbreviation in this study) Lumipulse HBsAg‐HQ Architect HBsAg QT HISCL HBsAg Lumipulse HBsAg ADVIA Centaur HBsAg II
Manufacturer Fujirebio Inc. Abbott Japan Co. Ltd. Sysmex Co. Ltd. Fujirebio Inc. Siemens Healthcare Diagnostics Inc.
Principle of operation CLEIA CLIA CLEIA CLEIA CLIA
Unit IU/mL (quantitative) IU/mL (quantitative) IU/mL (quantitative) COI (qualitative) COI (qualitative)
Capture antibodies Monoclonal (two) Monoclonal (two) Monoclonal (various) Polyclonal Monoclonal
Conjugate antibodies Monoclonal (two) Polyclonal Monoclonal (various) Monoclonal (two) Monoclonal
Reaction time (min) 30 30 17 30 30
Sample volume (μL) 100 75 20 100 100
Positive criterion ≥0.005 IU/mL ≥0.05 IU/mL ≥0.03 IU/mL COI ≥ 1.0 COI ≥ 1.0
Measuring range (theoretical value range) 0.005‐150 IU/mL (auto dilution) 0.05‐250 IU/mL (manual/auto dilution) 0.03‐2500 IU/mL (auto dilution) 0.1‐2000 COI 0.1‐1000 Index

CLEIA, Chemiluminescent Enzyme Immuno Assay; CLIA, Chemiluminescent Immuno Assay.