Skip to main content
. 2016 Nov 21;31(5):e22099. doi: 10.1002/jcla.22099

Table 5.

Proportion of laboratories by control rules or control frequency and percentages of acceptable laboratories against different allowable imprecision specifications for HbA1c (%)a

IQC practice Proportion of laboratories Allowable imprecision specifications based on NACB and biological variationb
2012 2013 2014 2015 2016 NACBc Minimum Desirable Optimal
IQC rules
Single IQC rule 18.3 18.5 10.9 7.9 7.6 30.8 21.7 5.6 0
12s/13s 4.0 3.3 2.5 8.2 9.3 50.8 40.2 19.5 5.1
13s/22s 10.8 14.4 14.2 20.0 19.0 60.1 39.8 18.6 0
12s/13s/22s 2.2 3.3 5.5 8.6 10.1 63.2 43.2 16.5 6.9
13s/22s/R4s 5.0 4.5 8.0 12.1 15.6 65.2 42.1 17.9 5.4
One or more from the rules of 12s, 13s, 22s, R4s, 41s and 10X¯ 15.1 18.1 16.7 37.5 32.9 59.5 39.9 13.2 6.5
Other rules 2.2 0.8 0.4 2.5 4.2 48.9 30.2 8.9 5.3
Unclear 42.4 37.0 41.8 3.2 1.3 35.6 20.5 6.8 0
IQC frequency
One QC run every more than 2 days 27.5 14.7 12.4 14.7 7.9 33.2 25.6 0 0
One QC run every 1‐2 days 71.9 84.5 85.9 82.4 57.7 61.2 39.4 18.9 5.2
More than one QC run per day 0.6 0.8 1.8 2.8 34.4 67.8 40.3 20.1 7.9
a

The subgroup analyses of the current CVs, mass fraction and pass rates were all for level 1.

b

The pass rates against four criteria were for the 2016 data.

c

Analytical imprecision criteria based on National Academy of Clinical Biochemistry (NACB) was set as <2% within laboratory.