Skip to main content
. 2018 Aug 26;33(2):e22659. doi: 10.1002/jcla.22659

Table 2.

Intra‐assay and inter‐assay precision

Group Low Medium High
IgM IgG IgM IgG IgM IgG
Intra‐assay precision (n = 20) Concentration (S/CO) 1.75 1.87 5.91 18.36 20.54 88.63
CV (%) 4.23 5.12 3.25 3.48 2.87 2.52
Inter‐assay precision (n = 10) Concentration (S/CO) 1.68 1.73 5.84 18.65 20.42 87.45
CV (%) 7.71 8.54 6.63 6.46 5.13 5.38