. 2019 Oct 23;10:1330. doi: 10.3389/fphys.2019.01330

TABLE 2.

Carotid stiffness parameters of BAV patients and controls.

BAV patientsN = 92 ControlsN = 48 p
PWV-1 (m.s–1) 4.40 ± 1.06 4.14 ± 1.07 0.188
PWV-2 (m.s–1) 6.42 ± 1.98 6.02 ± 2.05 0.293
Delta–PWV (m.s–1) 2.01 ± 1.87 1.88 ± 1.92 0.723
PWV-1/DBP (cm.s–1.mmHg–1) 6.82 ± 1.47 6.38 ± 1.18 0.084
PWV-2/SBP (cm.s–1.mmHg–1) 5.29 ± 1.63 5.24 ± 1.39 0.660
Delta–PWV/PP (cm.s–1.mmHg–1) 2.87 ± 3.08 3.38 ± 2.81 0.353
Distensibility (mmHg–1) 26.1 ± 21.0 27.6 ± 15.6 0.503
MRSD (s–1) 1.89 ± 1.10 2.15 ± 1.07 0.177
MRDR (s–1) 1.00 ± 0.59 1.01 ± 0.52 0.942

Results are mean ± standard deviation. BAV, bicuspid aortic valve; DBP, diastolic aortic pressure; SBP, systolic aortic pressure; PP, pulse pressure; PWV, pulse wave velocity; MRSD, maximal rate of systolic distension; MRDR, maximal rate of diastolic recoil. P-values are from Student t-test.