Skip to main content
. 2019 Oct 31;20(8):328. doi: 10.1208/s12249-019-1532-7

Table II.

Schott 6R Vial Heat Transfer Coefficient in REVO Based on Experimental Drying Time at Different Chamber Pressures

Chamber pressure (mTorr) Experimental drying time (h) Simulated drying time (h) Kv
W/m2 K 10−4 cal/s K cm2
100 12.82 12.81 15.06 3.6
300 11.62 11.62 21.34 5.1
1500 15.84 15.84 44.64 10.67