. 2019 Oct 28;25(40):6053–6062. doi: 10.3748/wjg.v25.i40.6053

Table 2.

Quantitative ultrasound: Optimal cutoffs obtained using liver biopsy as the reference standard

Technique S0 vs S1–S3 S0–S1 vs S2–S3 S0–S2 vs S3
CAP[24,25], dB/m (viral hepatitis) 219-222 233-296 290
Sensitivity: 76-91% Sensitivity: 60-87% Sensitivity: 78%
Specificity: 52-71% Specificity: 74-91% Specificity: 93%
CAP[26-28], dB/m (mixed etiologies) 255-266 283-311 293-318
Sensitivity: 65-75% Sensitivity: 57-86% Sensitivity: 83-88%
Specificity: 82-87% Specificity: 85-94% Specificity: 81-91%
CAP[29], dB/m1 (mixed etiologies) 248 268 280
Sensitivity: 69% Sensitivity: 77% Sensitivity: 88%
Specificity: 82% Specificity: 81% Specificity: 78%
CAP[32], dB/m (NAFLD) 302 331 337
Sensitivity: 80% Sensitivity: 70% Sensitivity: 72%
Specificity: 83% Specificity: 76% Specificity: 63%
ATI[41], dB/cm/MHz (mixed etiologies) 0.63 0.70 0.74
Sensitivity: 74% Sensitivity: 86% Sensitivity: 100%
Specificity: 77% Specificity: 81% Specificity: 82%
UGAP[42], dB/cm/MHz (mixed etiologies) 0.53 0.60 0.65
Sensitivity: 81% Sensitivity: 86% Sensitivity: 84%
Specificity: 87% Specificity: 81% Specificity: 90%
ATT[44], dB/cm/MHz (mixed etiologies) 0.62 0.67 0.73
Sensitivity: 72% Sensitivity: 87% Sensitivity: 82%
Specificity: 82% Specificity: 72% Specificity: 89%
1

Individual patient data meta-analysis. In parentheses: etiology of liver disease. CAP: Controlled attenuation parameter; dB: decibel; ATI: Attenuation imaging; MHz: Megahertz; UGAP: Ultrasound guided attenuation parameter; ATT: Attenuation.