Skip to main content
. 2019 Mar 12;15:100094. doi: 10.1016/j.jctube.2019.100094

Table 4.

Extrapulmonary specimens: comparison of Ultra vs. Xpert semiquantitative readouts with AFB smear and culture results.

No. of specimens with ultra Specimens with No. of specimens with Xpert
Not detected Smear positive growth of MTB (no. = 11) Not detected
Trace
Very low Very low
Low 1 2 Low
Medium 8 6 Medium
High 2 3 High
Total 11 11 Total
Not detected 2 Smear negative growth of MTB (no. = 40) 11 Not detected
Trace 5
Very low 14 6 Very low
Low 6 15 Low
Medium 13 6 Medium
High 2 High
Total 40 40 Total
Not detected 4 Smear positive growth of NTM (no. = 4) 4 Not detected
Trace
Very low Very low
Low Low
Medium Medium
High High
Total 4 4 Total
Invalid/Error 1 Smear negative no growth of AFB (no. = 35) 3 Invalid/Error
Not detected 33 32 Not detected
Trace 1
Very low Very low
Low Low
Medium Medium
High High
Total 35 35 Total