Table 1.

Selected Existing Technologies on CTC Enrichment

devices key observations (recovery rate, throughput, purity) reference
micropillar chip; herringebone chip 5–1281 CTCs/mL with 1 mL/h, detection rate 65%; 63 CTCs/mL with 1.2 mL/h, detection rate 91.8 ± 5% for PC3; 50.3 CTCs/mL with 8 mL/h, detection rate 98% SKBR3 M. Toner15,3537
micropillar chip 30 CTCs/mL with 1 mL/h, detection rate 97 ± 3% B.J. Kirby38
sinusoidal chip 53 CTCs/mL with 15 mL/h, detection rates are 98% MCF-7 and 82% MDA-MB-231 S. A Soper3942
microchannel array 10 CTCs/mL with 36 μL/min, recover rate >95%; using additional grooved surface can achieve over 90% capture rate with >84% purity; throughput 3.6 mL/h Z. Fan43,44
μHall: antibody labeled magnetic nanoparticles recovery rate of MDA-MB-468 99%; purity 100%; throughput 3.25 mL/h; analyzing ovarian 7.6 CTCs/mL H. Lee45
CellSearch system positive selection to enrich tumor cells from whole blood; the enriched and stained cells are scanned and imaged on CellTrackersAnalyzer II 8, 4649
microfilter 0–12.5 CTCs/mL with 90 mL/h, detection rate 90% M. Toner35,50
portable filter MCF-7, SK-BR-3:5 CTCs/mL at 0.5 psi constant pressure, detection rate >90% R.J. Cote,51
microfilters 1–3 CTCs/mL with 2 mL/h, detection rate enhanced with antibody coating: 95% MCF-7; 97% MDA-MB-231 C.T. Lim,52
microsieve filter with uniform pore structure undiluted whole blood with flow rate <2 mL/min; MCF-7 recover rate >80%; accurately detected CTC from eight patients M. Li53
3D parylene-C micronlter selection by both size and deformability; collect >85% viable spiked MCF-7 cells Y. Tai28
suspended microchannel resonator flow rate 45 μL/h; analyze 500 cells/h to reach 73% unhealthy cells; healthy blood cells passage time <10 ms; CTC passage time >10 ms S. Manalis5456
lab-on-a-disc: centrifugal microfluidic size selection MCF-7 61% capture rate; have the potential for high throughput and easy detection by directly mL; throughput >3 mL/min; sensitivity 95.9 ± 3.1% recovery rate; selectivity > 2.5 log WBCs depletion Y. Cho57,58
vortex capture CTC by fluidic vortex without having CTCs experience deformation; throughput 7.5 mL whole blood in 20 min; about 60–90 min on the system with no sample prep; no staining or any kind of affinity capture of the cells is required; efficiency of capture 60–80% D. Di Carlo31,32,5961