Table 4.

Sealing time (Primary and terminal experiment)

Vessel type thermoSEAL mode (TSM) BiClamp® mode (BCM) p value
No. mean ± STD (range) [mm] No. mean ± STD (range) [mm]
All vessels 97 3.5 ± 0.69 (2.5–5.9) 88 7.3 ± 1.3 (3.5–11) < 0.0001
Arteries 55 3.4 ± 0.65 (2.5–5.9) 51 7.4 ± 1.4 (3.5–11) < 0.0001
Veins 36 3.6 ± 0.78 (2.5–5.2) 27 7.4 ± 1.4 (4.7–10) < 0.0001
Renal arteries 17 3.5 ± 0.45 (2.8–4.4) 15 7.4 ± 1.4 (5.1–9.8) < 0.0001
Vascular bundles 6 3.7 ± 0.54 (3.0–4.4) 10 6.9 ± 0.87 (5.7–8.7) < 0.0001

6% (11/196) of the vessels seals with unintended multiple activations could not be evaluated