Table 2.

Adherence to the technical parameters of PI-RADS v2 in Turkey

Questions PI-RADS v2 recommendation Mean Min–max Adherence, n (%)
Pelvic coil, how many channels? ≥16 channel 16.8 2–64 52 (85.2)
Coronal T2 Should be obtained 55 (90.2)
Sagittal T2 Should be obtained 58 (95.1)
Precontrast axial T1 Should be obtained 59 (96.7)
Postcontrast T1 Should be obtained 53 (86.9)
One sequence covering aortic bifurcation Should be obtained 42 (68.9)

Axial T2: Slice thickness (mm) ≤3 3.25 2.5–4.5 42 (68.9)
Gap 0 0.42 0–2.5 20 (32.8)
Field of view (mm) 120–200 205 140–320 46 (75.4)
Frequency voxel size (mm) ≤0.4 0.67 0.31–1.05 6 (9.8)
Phase voxel size (mm) ≤0.7 0.81 0,31–1.34 25 (41)
NEX 2.5 1–5
Acquisition time (s) 233 76–490

Diffusion: Slice thickness (mm) ≤4 3.67 3–6 52 (85.2)
Gap 0 0.38 0–1.5 31 (50.8)
Field of view (mm) 160–220 255 140 – 461 23 (37.7)
Frequency voxel size (mm) ≤2.5 2.02 0.7–3.24 48 (78.7)
Phase voxel size (mm) ≤2.5 2.32 0.89–3.96 38 (62.3)
TR (ms) ≥3000 4876 400–8300 54 (88.5)
TE (ms) ≤90 80 55–116 47 (77)
Maximum b value (s/mm2) ≥1400 1302 600–2400 35 (57.4)
NEX of maximum b value 6.9 1–20
Acquisition time (s) 274 54–639

Dynamic*: Temporal resolution (s) ≤10, preferred <7 14.5 3.4–61 33 (55.9), 10 (16.9)
Slice thickness (mm) ≤3 3.34 0.9–4.8 23 (39)
Gap 0 0.4 0–3 30 (50.8)
Frequency voxel size (mm) ≤2 1.38 0.58–3.54 55 (93.2)
Phase voxel size (mm) ≤2 1.64 0.66–3.85 53 (89.9)
TR (ms) ≤100 30 2.7–500 56 (94.9)
TE (ms) ≤5 3.18 0.8–50 56 (94.9)
Acquisition time (s) ≥120 247 63–551 53 (89.9)

PI-RADS v2, Prostate Imaging-Reporting and Data System; min, minimum; max, maximum; NEX, number of excitations; TR, repetition time; TE, echo time.

*

Results of 59 clinics.