TABLE 1.

Summary of calibration and sample handling by instrument1

FiltMid FTMid1 FTMid2 Near
Operator Trained milk bank technician Trained milk bank technician Trained milk bank technician Trained milk bank technician
Calibration set Manufacturer calibration samples Human milk samples (10) analyzed by USDA Study calibration samples (10) Study calibration samples (10)
Volume required for duplicate readings 6 mL 50 mL 60 mL 6 mL
Thawing Overnight in refrigerator Overnight in refrigerator Overnight in refrigerator Overnight in refrigerator
Warming To 40°C in a bead warmer (heater model JBN5 US, Miris) Temperatures at analysis 4°–8°C. Pilot study on effect of temperature showed no difference in results at starting temperatures of 4°C, 24°C, and 40°C (unpublished data). To 20°C in water bath (Precision 183, Fisher Scientific) To 40°C in bead warmer (heater model JBN5 US, Miris)
Mixing Per manufacturer instructions Sonicated at 75% amplitude 2 min with 6-mm probe submerged ∼75% into sample (VCX 130, Sonics and Materials) Sonicated at 75% amplitude 3 min with 6-mm probe submerged ∼75% into sample (VCX 130, Sonics and Materials) Sonicated at 50% amplitude for 15 s (ultrasonic processor model CV18, Miris); sample gently inverted 10 times before loading
1

FiltMid (Miris HMA), FTMid1 (MilkoScan), FTMid2, (LactoScope), and Near (SpectraStar). FiltMid, filtered midinfrared; FTMid, Fourier-transformed full-spectra midinfrared; Near, near infrared.