Table 1.

Agreement between cone beam computed tomography defined radiological focus and film test (deviation vector is the average of the three‐dimensional vector of each independent measurement).

Coordinate ∆X ∆Y ∆Z
100, 100, 100 −0.07 ± 0.03 mm 0.02 ± 0.17 mm −0.1 ± 0.17 mm
Deviation vector 0.29 ± 0.06 mm