Table 4.

Difference between fiducial localization and cone beam computed tomography (CBCT) localization on the Ball phantom.

CTDI 2.5 CTDI 6.3
  Rotation (°) Translation (mm)   Rotation (°) Translation (mm)
X Y Z X Y Z X Y Z X Y Z
CBCT 1 0.46 −0.12 0.41 −0.18 −0.34 −0.94 CBCT 1 0.51 −0.02 −0.08 −0.13 −0.19 −0.64
CBCT 2 0.64 0.11 −0.03 −0.13 −0.27 −0.7 CBCT 2 0.24 0.08 −0.26 −0.17 −0.26 −0.54
CBCT 3 0.13 −0.31 0.02 0.08 −0.04 −0.81 CBCT 3 0.51 0.15 0.07 −0.12 −0.28 −0.79
Average 0.41 −0.11 0.13 −0.08 −0.22 −0.82 Average 0.42 0.07 −0.09 −0.14 −0.24 −0.66