Table 1.
Instrument | Tilt scheme | Unblank timea | Prep timeb | Set Rec timec | Cam Rec timed | Total timee |
---|---|---|---|---|---|---|
[ST/K3] | unidirectional | 1 s | 98 s | 91 s | 95 s | 201 s |
[DT/K2] | unidirectional | 1 s | 136 s | 117 s | 253 s | 399 s |
[ST/K3] | dose-symmetric | 1 s | 101 s | 130 s | 134 s | 243 s |
[DT/K2] | dose-symmetric | 1 s | 134 s | 185 s | 339 s | 484 s |
Time the sample is exposed per tilt.
Time taken to refine eucentricity and autofocus.
Exposure time set for Record in SerialEM. Variations are due to the slower tilt speed on Titan Krios [DT/K2]. In all cases, the default tilt speed of the stage was not altered.
Time from start of acquisition until regain of microscope control.
Total run time of the tilt series including resetting the microscope state.