Skip to main content
. 2019 Oct 15;7(10):e2466. doi: 10.1097/GOX.0000000000002466

Table 1.

Average Surface Metrology Parameters of Silicone Breast Implant Surfaces Measured According to ISO 14607:2018

Parameter (±SD) Traditional Smooth (Allergan) SmoothSilk/SilkSurface VelvetSurface Siltex Sientra and Silimed Microcell BIOCELL
Roughness (Ra) (µm) 0.3 ± 0.2 3.2 ± 0.6 14.6 ± 2.5 32.0 ± 5.0 40.0 ± 6.0 53.0 ± 9.0 79.0 ± 14.0
Density of peaks (peaks/cm2) 5,996 ± 6,105 25,820 ± 5,766 2,897 ± 980 2,008 ± 1,605 NA NA NA
Ku (Sku) 5.0 ± 2.0 4.1 ± 0.6 3.0 ± 0.7 2.9 ± 0.3 3.1 ± 0.4 2.0 ± 0.1 3.0 ± 1.0
SA (mm2) 1.0 ± 0.0 1.15 ± 0.04 1.16 ± 0.02 1.87 ± 0.29 1.5 ± 0.2 1.8 ± 0.2 2.92 ± 0.52
ISO 14607:2018 Surface classification Smooth Smooth Microtextured Microtextured Microtextured Macrotextured Macrotextured

Density of peak measurements cannot be performed for BIOCELL, Microcell, and Sientra/Silimed surfaces (NA). Results property of Establishment Labs.