Table 2.

Comparing the differences in the reference ranges of TEG parameters determined in this study and provided by manufacturers in evaluating neonatal coagulation

Parameter This study Manufacturer provided N/N N/A A/N A/A
R 2.80‐7.17** 5.00‐10.00 162 195 4 10
Angle 44.73‐78.88** 53.00‐72.00 199 155 0 17
K 0.80‐4.57** 1.00‐3.00 277 87 0 7
MA 43.92‐74.65** 50.00‐70.00 262 91 0 18
LY30 0.10‐6.94 0.00‐8.00 363 0 2 6

The McNemar test was used for the analysis between the reference ranges determined in our study and provided by manufacturers. R (s), K (s), Angle (α), maximum amplitude (MA, mm) and lysis at 30 min (LY30, %).

Abbreviation(s): A/A, abnormal data evaluated by both methods; A/N, abnormal data evaluated by this study, normal data evaluated by manufacturers; N/A, normal data evaluated by this study, abnormal data evaluated by manufacturers; N/N, normal data evaluated by both methods; TEG, thromboelastograph.

**

P < 0.01.