Table 1. Quality indicators implemented to monitor the performance of Xpert MTB/RIF molecular assay in Ethiopia.

Indicator Numerator Denominator Target
Indicator 1: Percentage of specimens reported as MTB detected (MTB positivity rate) Number of specimens reported as MTB detected RIF resistance not detected, RIF resistance detected, and RIF indeterminate in 1 month Total number of specimens tested in 1 month 13.42–24.61%[1318]
Indicator 2: Percentage of specimens reported as MTB detected; rifampicin resistance detected (RIF resistance rate) Number of specimens reported as MTB detected RIF resistance detected in 1 month Total number of MTB detected in 1 month 5.8% (2.8–8.4%)[12]
Indicator 3: Percentage of specimens reported as MTB detected; rifampicin indeterminate (RIF resistance indeterminate rate) Number of specimens reported as MTB detected RIF indeterminate in 1 month Total number of MTB detected in 1 month 8.9%[17] 13.6%[13]
Indicator 4: Percentage of specimens with error results (Error rate) Number of specimens with error results in 1 month Total number of specimens tested in 1 month <3%[11]
Indicator 5: Percentage of specimens with invalid results (Invalid rate Number of specimens with invalid results in 1 month Total number of specimens tested in 1 month <1%[11]
Indicator 6: Percentage of specimens with “no results” (No result rate) Number of specimens with “no results” in 1 month Total number of specimens tested in 1 month <1%[11]
Indicator 7: Percentage of Xpert MTB/RIF results reported within TAT for results (Within TAT rate) Number Xpert MTB/RIF results reported within the target TAT (2–24hrs) for results Total number of Xpert MTB/RIF results reported 90%[11]