Table 1.

Syringe volume errors produced by SP, WA, and PolyLS methods of pneumotachograph calibration

Entire Data Set
Testing Data Set
Method Average, % CI95% Average, % CI95%
SP 0.29* –0.39 to 1.00 0.29* –0.46 to 0.97
WA 0.03 –0.88 to 0.62 0.03 –1.02 to 0.61
PolyLS 0.01 –1.04 to 0.47 0.01 –1.15 to 0.50

Volume errors are expressed as % of 3 liters. CI95%, 95th percentile confidence interval of volume errors; PolyLS, polynomial least squares; SP, syringe potentiometer; WA, weighted averaging. Note: The CI95% of volume errors for the 3 methods all lie within the American Thoracic Society/European Respiratory Society-recommended error margin of ±3.50%.

*

Significantly different from WA method, P < 0.05.

Significantly different from PolyLS method, P < 0.05.