. 2019 Nov-Dec;36(6):567–568. doi: 10.4103/lungindia.lungindia_96_19

Use of Venturi to prevent desaturation during nebulization

Wasimul Hoda 1, Brajesh Kumar Ratre 2, Sachidanand Jee Bharti 3
PMCID: PMC6852213  PMID: 31670313

Sir,

Nebulization is the delivery of saline or drugs in the form of small droplets to the distal lung, which is a common practice in intensive care unit (ICU).[1] It is done with the help of nebulizers which are connected to the inspiratory limb of breathing circuit either in mechanically or spontaneously ventilated patients. Spontaneously breathing patients who are admitted in ICU also need oxygen supplementation. Nebulization in such patients needs attention with careful monitoring of vitals, airway pressure, and oxygen saturation to prevent any desaturation or other adverse events.[2] During conventional nebulization, even with oxygen flow rate of 6–8 L/min, there is a fall in oxygen saturation. In some cases, if not corrected, it may lead to severe hypoxemia and hemodynamic instability. Therefore, we introduce a noble technique using a Venturi to prevent such events. Venturi is considered a high-flow oxygen therapy device. The final assembly with Venturi is shown in Figure 1. It is connected between the nebulizer and the oxygen source. Different types of Venturi device (color coded) can be used according to the need. The assembly increases the flow, which helps in the delivery of high FiO2 preventing desaturation.

Figure 1.

Figure 1

Venturi–nebulizer assembly; Venturi is attached between the nebulizer apparatus and oxygen source

In literature, there are numerous other advantages of high-flow therapy. Venturi is commonly used in the hospital setting to deliver controlled percentages of oxygen.[3] During acute distress, the ventilatory demand increases to 30 L/min to 120 L/min. Normal quite breathing inspiratory flow is 15 L/min. Venturi nebulizer creates inspiratory flow of 60 L/min. By increasing inspiratory flow demand, it also gives a subjective feeling of satisfaction with an objective decrease in tachypnea and thereby improves saturation.[4] As it is a high-flow delivery device, it creates positive pressure in nasopharynx which could act like positive end expirarory pressure (PEEP) to prevent alveolar collapse at the end of expiration. Hence, it helps improve ventilation and perfusion ratio, minimizing shunting.[5]

CONCLUSION

In ICU settings, patients are oxygen dependent due to intrinsic lung pathology or because of weakness in respiratory musculature. Many a time, desaturation may lead to suboptimal respiratory care and premature cessation of nebulization. This technique of using Venturi system with nebulization effectively takes care of the desaturation and improves care in ICU.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

Acknowledgments

We would like to thank the ICU team along with the physiotherapist colleague working with us during the management of such cases.

REFERENCES

  • 1.O'Doherty MJ, Thomas SH. Nebuliser therapy in the intensive care unit. Thorax. 1997;52(Suppl 2):S56–9. doi: 10.1136/thx.52.2008.s56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Dhanani J, Fraser JF, Chan HK, Rello J, Cohen J, Roberts JA, et al. Fundamentals of aerosol therapy in critical care. Crit Care. 2016;20:269. doi: 10.1186/s13054-016-1448-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Pepperell JC, Fraser C, Mcclue G. The use of Venturi masks with oxygen concentrators. Thorax. 2011;201054c:139. [Google Scholar]
  • 4.L'Her E, Deye N, Lellouche F, Taille S, Demoule A, Fraticelli A, et al. Physiologic effects of noninvasive ventilation during acute lung injury. Am J Respir Crit Care Med. 2005;172:1112–8. doi: 10.1164/rccm.200402-226OC. [DOI] [PubMed] [Google Scholar]
  • 5.Parke R, McGuinness S, Eccleston M. Nasal high-flow therapy delivers low level positive airway pressure. Br J Anaesth. 2009;103:886–90. doi: 10.1093/bja/aep280. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Lung India : Official Organ of Indian Chest Society are provided here courtesy of Wolters Kluwer -- Medknow Publications

RESOURCES