Skip to main content
. 2019 Aug 23;83(1):228–239. doi: 10.1002/mrm.27926

Table 1.

Scan parameters

Phantom Mouse
Intramuscular Intravascular
19F 19F 1H 19F 1H
Sequence 3D FLASH 3D FLASH 3D FLASH 3D FLASH 3D FLASH
Readout direction FH, HF, LR, RL FH and LR FH FH and LR FH
Acquisition matrix 128 × 128 × 128 128 × 128 × 128 128 × 128 × 128 64 × 64 × 64 128 × 128 × 128
Field of view (mm3) 40 × 40 × 40 40 × 40 × 40 40 × 40 × 40 40 × 40 × 40 40 × 40 × 40
TE (ms) 3.03 3.03 2.76 2.76 2.76
TR (ms) 20 20 20 20 20
Flip angle 35° 25° 25° 25° 25°
Excitation BW (Hz) 33,000 33,000 33,000 33,000 33,000
Receiver BW (Hz) 44,600 44,600 44,600 44,600 44,600
No. of averages 4 6 2 24 4
Scan time 4 × 21m51s 2 × 32m46s 10m55s 2 × 32m46s 10m55s

BW = bandwidth; FH = foot‐to‐head; HF = head‐to‐foot; LR = left‐to‐right; RL = right‐to‐left.