. 2019 Aug 1;46(9):3893–3905. doi: 10.1002/mp.13696

Table 1.

Sequence parameters

  T2W DWI 31P MRSIb 1H MRSI
Sequence type Turbo‐Spin‐Echo EPI (RESOLVE); b0, b100, b400, and b800 3D MRSI with non‐selective BIR‐4 excitation PRESS‐like with spectrally and spatially selective RF refocusing pulses
FOV 111 × 111 mm2 154 × 154 mm2

a): 120 × 100 × 100 mm3

b): 120 × 120 × 120 mm3

84 × 70 × 70 mm3
Matrix 256 × 256 88 × 88

a): 12 × 10 × 10

b): 10 × 10 × 10

12 × 10 × 10
Voxel sizea 0.43 × 0.43 × 3 mm3 1.75 × 1.75 × 3 mm3

a): 4.9 cc

b): 9.1 cc

1.4 cc
Slices 19 19 n.a. n.a.
Slice thickness 3 3 n.a. n.a.
TR (ms) 4250 4400 1500 1000
TE (ms) 84 68 n.a. 135
Bandwidth (Hz) 106 (per pixel) 1015 10000 (spectral) 2400 (spectral)
Vector size n.a. n.a. 2048 1024
Averages 1 1 4 (weighted) 1
Acquisition time (min:sec) 1:59 4:34 13:09 7:01
Flip angle (⁰) n.a. n.a. 45 n.a.
Filter Prescan normalization filter n.a. 100% Hamming filter 50% Hamming filter

DWI, diffusion‐weighted imaging; MRSI, magnetic resonance spectroscopic imaging.

a

For the MRSI measurements, this was based on the true voxel size based on voxel volume within the 50% range of the maximum of the point spread function after correcting for spatial filter and sampling scheme.

b

Two different measurement parameter sets were used: set a) and set b).