Table 4.

Normal range definition of the STG‐DrugScreen application on the ST Genesia system using plasma sample from 42 healthy individuals

Healthy (n = 42) Lag time absolute Peak height absolute Time‐to‐peak absolute ETP absolute Lag time normalized (ratio) Peak height normalized Time‐to‐peak normalized (ratio) ETP normalized
Mean 0.93 min 456.20 nmol/L 2.03 min 1418 nmol/L 1.16 97% 0.98 83%
SD 0.09 min 56.35 nmol/L 0.16 min 270 nmol/L 0.10 12% 0.08 15%
CV (%) 9.8 12.4 7.9 19.1 9.6 12.1 7.9 18
Minimum 0.73 min 355.90 nmol/L 1.80 min 1027 nmol/L 0.81 75% 0.87 56%
Maximum 1.22 min 571.40 nmol/L 2.69 min 1993 nmol/L 1.33 120% 1.29 118%
10% 0.82 min 380.20 nmol/L 1.87 min 1122 nmol/L 0.90 81% 0.90 68%
90% 1.00 min 538.90 nmol/L 2.17 min 1851 nmol/L 1.11 115% 1.29 109%

All samples were tested on the same day, explaining the few differences in term of variability between absolute and normalized results.

Abbreviations: CV, coefficient of variation; ETP, endogenous thrombin potential; SD, standard deviation.