Comparison of the miniature spectrometers QE65 Pro and customized Ventana VIS-NIR.

Specifications QE65 Pro Customized Ventana
Light throughput F/4 F/2
Bandwidth 353 nm (700 to 1053 nm) 677 nm (430 to 1100 nm)
Spectral resolution 3.5 nm 4 nm
CCD peak quantum efficiency 95% 75%
CCD pixel size 24.6  μm2 14  μm2
CCD pixels 1024×58 1024×64
CCD active area 24.576×1.392mm2 14.336×0.896mm2
CCD temperature 15°C (40°C below ambient) 15°C (10°C below ambient)
Dark noise (single spectrum) 3 counts RMS 6 counts RMS
SNR 1000:1 500:1