Figure 5.

Figure 5.

The mean and standard deviations of the parameter values (DLC/d, coronary artery height (h) and Sinus of Valsalva diameter (SOVd)) for those high risk patients who successfully received a transcatheter aortic valve replacement without coronary obstruction are compared to those who did not receive a transcatheter aortic valve replacement successfully. A significant difference between the two groups was found for the DLC/d parameter at significance level 0.05. Neither h nor SOVd was significantly different between the two groups.