TABLE 3.

Characteristics of phenotypic tests for carbapenemase detectiona

Test method Specific detection of OXA-48 Sensitivity/specificity (%) Sensitivity for OXA-48 (%) Com Special equipment Interpretation Exper TAT Cost (U.S. dollars) Ref
Inhibitor-based, MASTDISCS Combi Carba plus Indirectly 100/91.1 100 Yes No Disk zone diam Low 18–24 h 234
Carba NP test, RAPIDEC CARBA NP No 97.8–99/98.5–100 93–96.3 Yes No Color change Low 2.5 h $10 244
Colorimetric, β-CARBA test No 85.1/92.7 80.5b Yes No Color change Low 35 minb 258
mCIM, CLSI No 97/99 100 No No Disk zone diam Low 22 h $1 244
rCIM No 99/100 97.4 No No Disk zone diam Low 2.5 h $0.25 260
Lateral flow immunoassay, O.O.K. K-SeT Yesc OXA-48 and KPC only 100 Yes No Visible lines Low 20 min $11
Lateral flow immunoassay, NG-Test CARBA5 Yes 97.3/99.8 100 Yes No Visible lines Low 20 min 268
Electrochemical assay, BYG Carba v2.0 No 96.3/99.7 97.2 No Homemade potentiostat Conductance signal with cutoff value High 30 min $2.5 270
Spectrophotometry No 100/98.5 100 No UV spectrophotometer Absorbance slope with cutoff value High 12–24 h $2–3 271
MALDI-TOF MS, MBT STAR-Carba IVD No 100/98.2 100 Yes MALDI-TOF MS (Bruker) Automatic interpretation by software Low 55 min ? 280
a

Com, commercially available; Exper, expertise needed; TAT, turnaround time; Ref, reference.

b

A higher sensitivity for OXA-48 producers was reported when an extended incubation time (60 min) was used.

c

OXA-48 groups and OXA-163 groups can be separately detected.