TABLE 3.
Test method | Specific detection of OXA-48 | Sensitivity/specificity (%) | Sensitivity for OXA-48 (%) | Com | Special equipment | Interpretation | Exper | TAT | Cost (U.S. dollars) | Ref |
---|---|---|---|---|---|---|---|---|---|---|
Inhibitor-based, MASTDISCS Combi Carba plus | Indirectly | 100/91.1 | 100 | Yes | No | Disk zone diam | Low | 18–24 h | 234 | |
Carba NP test, RAPIDEC CARBA NP | No | 97.8–99/98.5–100 | 93–96.3 | Yes | No | Color change | Low | 2.5 h | $10 | 244 |
Colorimetric, β-CARBA test | No | 85.1/92.7 | 80.5b | Yes | No | Color change | Low | 35 minb | 258 | |
mCIM, CLSI | No | 97/99 | 100 | No | No | Disk zone diam | Low | 22 h | $1 | 244 |
rCIM | No | 99/100 | 97.4 | No | No | Disk zone diam | Low | 2.5 h | $0.25 | 260 |
Lateral flow immunoassay, O.O.K. K-SeT | Yesc | OXA-48 and KPC only | 100 | Yes | No | Visible lines | Low | 20 min | $11 | |
Lateral flow immunoassay, NG-Test CARBA5 | Yes | 97.3/99.8 | 100 | Yes | No | Visible lines | Low | 20 min | 268 | |
Electrochemical assay, BYG Carba v2.0 | No | 96.3/99.7 | 97.2 | No | Homemade potentiostat | Conductance signal with cutoff value | High | 30 min | $2.5 | 270 |
Spectrophotometry | No | 100/98.5 | 100 | No | UV spectrophotometer | Absorbance slope with cutoff value | High | 12–24 h | $2–3 | 271 |
MALDI-TOF MS, MBT STAR-Carba IVD | No | 100/98.2 | 100 | Yes | MALDI-TOF MS (Bruker) | Automatic interpretation by software | Low | 55 min | ? | 280 |
Com, commercially available; Exper, expertise needed; TAT, turnaround time; Ref, reference.
A higher sensitivity for OXA-48 producers was reported when an extended incubation time (60 min) was used.
OXA-48 groups and OXA-163 groups can be separately detected.