Skip to main content
. 2019 Nov 12;7:322. doi: 10.3389/fbioe.2019.00322

Table 2.

Measurement differences between the outREACH sensor and Vicon systems.

Coefficient of variation % Mean difference Lower limit of agreement Upper limit of agreement
ARAT block
Movement time 3.30 −16.00 −200.00 170.00
Spectral arc length −2.40 0.00 −0.11 0.11
Peak velocity 3.20 18.00 −55.00 92.00
Drink glass
Movement time 2.70 5.00 −220.00 230.00
Spectral arc length −8.50 −0.07 −0.53 0.38
Peak velocity 8.40 −82.00 −220.00 51.00
Pour water
Movement time 3.20 11.00 −290.00 310.00
Spectral arc length −7.70 0.10 −0.32 0.53
Peak velocity 13.00 −45.00 −230.00 140.00