Table 2.
Measurement differences between the outREACH sensor and Vicon systems.
| Coefficient of variation % | Mean difference | Lower limit of agreement | Upper limit of agreement | |
|---|---|---|---|---|
| ARAT block | ||||
| Movement time | 3.30 | −16.00 | −200.00 | 170.00 |
| Spectral arc length | −2.40 | 0.00 | −0.11 | 0.11 |
| Peak velocity | 3.20 | 18.00 | −55.00 | 92.00 |
| Drink glass | ||||
| Movement time | 2.70 | 5.00 | −220.00 | 230.00 |
| Spectral arc length | −8.50 | −0.07 | −0.53 | 0.38 |
| Peak velocity | 8.40 | −82.00 | −220.00 | 51.00 |
| Pour water | ||||
| Movement time | 3.20 | 11.00 | −290.00 | 310.00 |
| Spectral arc length | −7.70 | 0.10 | −0.32 | 0.53 |
| Peak velocity | 13.00 | −45.00 | −230.00 | 140.00 |