Skip to main content
. 2019 Nov 14;12:461–467. doi: 10.2147/MDER.S206269

Table 2.

Reported Device-Related And Patient-Related Events By Timing In Relation To Index Procedure

Total During Procedure Early (<30 Days) Late (>30 Days) Unspecified Timing
N N (%) N (%) N (%) N (%)
Endoleak 169 19 (11.2%) 31 (18.3%) 51 (30.2%) 68 (40.2%)
Dissection, penetrating aortic ulcer and intramural hematoma 67 16 (23.9%) 13 (19.4%) 22 (32.8%) 16 (23.9%)
Retrograde dissection 35 6 (17.1%) 11 (31.4%) 12 (34.3%) 6 (17.1%)
Neurologic 54 3 (5.6%) 38 (70.4%) 1 (1.9%) 12 (22.2%)
Stroke 15 0 13 (86.7%) 1 (6.7%) 1 (6.7%)
Paraplegia 37 3 (8.1%) 24 (64.9%) 0 10 (27.0%)
Unspecified 2 0 1 (50%) 0 1 (50%)
Technical device failure 45 45 (100%) 0 0 0
Infections 18 0 3 (16.7%) 8 (44.4%) 7 (38.9%)
Aorto-esophageal and aorto-bronchial fistula 11 1 (9.1%) 2 (18.2%) 6 (54.5%) 2 (18.2%)
Migration 7 4 (57.1%) 0 0 3 (42.9%)