Table 2.
Reported Device-Related And Patient-Related Events By Timing In Relation To Index Procedure
Total | During Procedure | Early (<30 Days) | Late (>30 Days) | Unspecified Timing | |
---|---|---|---|---|---|
N | N (%) | N (%) | N (%) | N (%) | |
Endoleak | 169 | 19 (11.2%) | 31 (18.3%) | 51 (30.2%) | 68 (40.2%) |
Dissection, penetrating aortic ulcer and intramural hematoma | 67 | 16 (23.9%) | 13 (19.4%) | 22 (32.8%) | 16 (23.9%) |
Retrograde dissection | 35 | 6 (17.1%) | 11 (31.4%) | 12 (34.3%) | 6 (17.1%) |
Neurologic | 54 | 3 (5.6%) | 38 (70.4%) | 1 (1.9%) | 12 (22.2%) |
Stroke | 15 | 0 | 13 (86.7%) | 1 (6.7%) | 1 (6.7%) |
Paraplegia | 37 | 3 (8.1%) | 24 (64.9%) | 0 | 10 (27.0%) |
Unspecified | 2 | 0 | 1 (50%) | 0 | 1 (50%) |
Technical device failure | 45 | 45 (100%) | 0 | 0 | 0 |
Infections | 18 | 0 | 3 (16.7%) | 8 (44.4%) | 7 (38.9%) |
Aorto-esophageal and aorto-bronchial fistula | 11 | 1 (9.1%) | 2 (18.2%) | 6 (54.5%) | 2 (18.2%) |
Migration | 7 | 4 (57.1%) | 0 | 0 | 3 (42.9%) |