Skip to main content
. 2019 Nov 19;19:217. doi: 10.1186/s12890-019-0980-x

Table 2.

Trigger results for shallow nasal breathing, proportion of pulses aligned with inhalation

POC Setting Proportion of POC pulses aligned with inspiration
Device A Device B Device C
1 42% 100% 99%
2 40% 100% 90%
3 42% 100% 51%
4 44% 100% 78%