Electrochemical Methods | |||||||
---|---|---|---|---|---|---|---|
Target Analyte |
Detection Principle |
Sensing Element |
Sensing Material |
Substrate | LOD | Real Sample | Ref. |
Hg+2 | ASV | Electrode system | WE: Au | PDMS | 3 ppb | No | [101] |
CE: Au | |||||||
RE: Ag | |||||||
Pb+2 and Hg+2 |
CV | Electrode system | WE: carbon | Polyethylene terephthalate |
50 M each | No | [104] |
CE: Ag | |||||||
RE: Ag/Cl | |||||||
Pb+2 | SWASV | Electrode system | WE: Ag | Polymer | 0.55 ppb | No | [99] |
CE/QRE: Ag | |||||||
Pb+2 and Hg+2 |
SWASV | Electrode system | WE: carbon | Paper | 2.0 and 2.3 ppb, resp. water |
Soda water and dirty ground |
[111] |
CE: carbon | |||||||
RE: Ag pseudo | |||||||
As | CV | Electrode system | WE: Ag | Plastic | 1 ppb | No | [102] |
CE: Carbon | |||||||
RE: Ag/AgCl | |||||||
As | SWASV | Electrode system | WE: Au/SWCNT | PDMS | 4.5 ppb | No | [103] |
CE and RE: SWCNT | |||||||
As(III) | Amperometry | Bioreporter with electrode system |
WE: Au | Plastic | 0.8 ppb | Tap and ground water |
[112] |
CE and RE: Ag and E. coli | |||||||
Nitrate | CV | Electrode system | WE and RE: Ag | Glass | 25 ppb | [113] | |
CE: Au | |||||||
Nitrate | CV | Electrode system | WE: Ag | Glass | 0.2 ppm | Field and environmental water |
[114] |
CE: Au | |||||||
RE: Ag | |||||||
Nitrate and Potassium |
Potentiometric | Electrode system | WE: Polymeric membrane | Green tapes | 9.56 and 0.81 mg/L, respectively |
Water from recycling unit |
[115] |
RE: Ag/AgCl | |||||||
Nitrate | Potentiometric | Electrode system with modified working electrode | WE: f-MWCNTs RE:Ag/AgCl |
Lipophilic carbon nanotubes | −7 M | Desalinated seawater | [116] |
Nitrate | EIS | Electrode system with modified working electrode | WE: NiR/nTiO2-GF CE: Au RE: Ag/AgCl |
PDMS | 1 M | No | [117] |
Nitrate and Nitrite |
SWV | Electrode system with modified working electrode | WE: Cu/MWCNT/RGO/GCE CE: Pt-wire RE: Ag/AgCl |
GCE | 20 and 30 nM, resp. | Tap and mineral water | [77] |
E. coli | Voltammetry | Electrode system with modified working electrode | WE: immobilized DNA probe on Au CE and RE: Pt |
Glass | 100 nM | No | [118] |
Hepatitis B | ASV | Electrode system | WE: GCE CE: Pt wire RE: Ag/AgCl and DNA modified AgNP |
Paper | 85 pM | No | [119] |
E. coli | Amperometry | Immunoassays | Antibody | PMMA | 50 CFU/mL | Real sample | [120] |
E.coli | Positive dielectrophoresis |
Sensing and focusing electrode | Not specified | PDMS | 300 CFU/mL | No | [121] |
E. coli | EIS | Interdigitated electrodes | Modified silicon sensor chip |
Silicon and PDMS |
10 cells/mL | No | [122] |
E. coli and S. aureus |
EIS | Coplanar electrode | Au electrode | Silicon | 522 cells/mL | No | [123] |
E. coli and Enterococci |
Coulter principle |
Microfluidic sensing chip |
Resistance detection circuit |
PDMS | Individual cell | Ballast water sample |
[124] |
Salmonella typhimurium |
Impedance analyzer | Interdigitated electrodes |
Au electrode with immobilized antibodies |
PDMS | 3 CFU/mL | No | [125] |
LOD: Limit of detection; ASV: Anodic stripping voltammetry; PDMS: Polydimethylsiloxane; PMMA: Poly(methyl methacrylate); CV: Cyclic voltammetry; SWASV: Square-wave anodic stripping voltammetry; QRE: Quasi-reference electrode; WE:Working electrode; CE: Counter electrode; RE: Reference electrode; EIS: Electrochemical impedance spectroscopy; SWV: Square-wave voltammetry; RGO: Reduced graphene oxide; GCE: Glassy carbon electrode; -MWCNTs: Functionalised- multiwall carbon nanotube.