Fig. 10.
Circular patch probe and interdigitated capacitor probe array with quadrifilar strip helix and longitudinal end strips: coronal (left) and sagittal (right) (a) simulated efficiency, (b) measured GRE images, and (c) measured efficiency inside of cylindrical phantom. Phasing shown in Table IV and magnitude excitations a quarter of the magnitude of the circular patch. Geometric distortions (barreling aspect) of are due to uncorrected non-linearity of the gradient coils.
