Table 2.

Comparative summary of DCT specifications (Kerr & Tupper, 2017)

Technology Detect a wide variety of compounds Ability to detect fentanyl and other opioids Ability to detect multiple compounds at once Specificity Sensitivity Quantitative analysis Identify unknown compounds Speed per sample
Colorimetric Reagent Testing Moderate Low Low Low Low No No <6 min
Fourier-Transform Infrared Spectroscopy (FTIR) High Moderate High High High Low No <2 min
Thin Layer Chromatography (TLC) Moderate Weak Moderate Moderate Moderate Low No 30 min
Capillary Electrophoresis with UV detection High Moderate Moderate Moderate Moderate Moderate No <2 min*
High Performance Liquid Chromatography with UV detection High High High High High High No 15 min
High Performance Liquid Chromatography with MS detection Highest Very high Very high Very high Highest Highest Yes 7.5 min*
*

These durations are estimates based on machine-specific run times alone, and do not include collection, preparation, report generation, or consultation1