Table IV.

Proposed cut-off values of radiological JSW in millimeters based on the measurement error (e.g., Bland and Altman method with SDD) in hip OA

Author Year Analysis* Film Bland and Altman method
N Mean change SD change SDD mm (cut-off) Limits of agreement
Min Max
Dougados 1996 3(52) 1 30 0.01 0.28 0.56 −0.55 0.57
Auleley 1998 1 1 (standing) 46 0.02 0.11 0.22 −0.2 0.24
1 1 (supine) 46 −0.002 0.13 0.26 −0.262 0.258
Auleley 2001 2 1 37 −0.02 0.23 0.46 −0.48 0.44
Conrozier 2001 1 1 28 0 0.15 0.3 −0.3 0.3
Maillefert 2002 3 (156) 1 35 0.028 0.39 0.78 −0.752 0.808
3 (156) 1 (computer) 35 0.077 0.34 0.67 −0.593 0.747
Hilliquin 2002 1 1 100 −0.03 0.22 0.44 −0.47 0.41
Maheu 2005 3 (156) 1 50 0.02 0.18 0.37 −0.35 0.39
3 (156) 2 50 0.03 0.16 0.31 −0.28 0.34
3 (156) 3 50 0.03 0.16 0.31 −0.28 0.34
*

Analysis: 1 = score of a single film in a single patient; 2 = score of two films in a single patient taken less than 2 days a part; 3 = score of the changes in a single patient between two visits and () the number of weeks between these two visits.

Film: 1 = pelvic; 2 = hip AP; 3 = hip lateral.