Target point dose comparison between the TPS and TLDs for all irradiation trials.

TLD location
TPS dose calculated, cGy (RBE)
Dose measured, cGy (RBE)
Measured-to-calculated dose ratio
Trial 1
 Target superior 646.2 654.3 1.013
 Target inferior 648.6 627.9 0.968
Trial 2
 Target superior 646.2 631.6 0.977
 Target inferior 648.6 638.1 0.984
Trial 3
 Target superior 646.2 631.8 0.978
 Target inferior 648.6 625.6 0.964
Trial 4
 Target superior 646.2 634.8 0.982
 Target inferior 648.6 646.3 0.996