Skip to main content
. 2018 Mar 21;4(3):40–47. doi: 10.14338/IJPT-17-00005.1

Table 3.

Two-dimensional gamma analysis pass rates for all irradiation trials.

Location
2-Dimensional gamma percentage of pixels passing
±5%/3 mm
±5%/4 mm
±7%/4 mm
Trial 1
 Axial 86.3% 91.0% 95.5%
 Sagittal 82.6% 87.6% 94.2%
Trial 2
 Axial 91.0% 94.0% 97.3%
 Sagittal 80.2% 87.0% 93.2%
Trial 3
 Axial 82.1% 88.6% 93.4%
 Sagittal 76.0% 81.8% 90.0%
Trial 4
 Axial 91.0% 93.4% 96.2%
 Sagittal 80.6% 86.4% 92.7%