Skip to main content

Table I.

Artifact measurements in phantom

Component description Material/design Artifact size (mm)
Electrodes
 Ag/AgCl: SLE UK; 131/9/TP cast pure Ag coated in AgCl; d: 9 mm 8/8
 Au: Grass USA; E5GH cast pure Ag with heavy Au coating; d: 10 mm 6/4
 Plastic Ag/AgCl: Meditec Italy; 1183 AGCL plastic coated in Ag/AgCl; d: 10 mm 4/4
 Carbon: Telefactor USA; carbon electrodes carbon electrodes with carbon wire; d: 10 mm 2/4
Resistors
 Carbon composition: Vitrohm Germany; series BT, 104‐0 carbon black, phenol resin, bakelite; length: 9.9 mm, d: 3.5 mm 19/19
 Planar thick film cermet, BI Technology, BPCE ruthenium oxide on aliminia substrate; 25.4/10.2/2.5 mm 6/4
 Cermet film: Meggitt CGS; HB01 cermet film on ceramic substrate with epoxy coating; 26.5/10.5/3.0 mm 2/2
Resistor insulation
 Epoxy rapid setting two part epoxy; 0.1 ml 8/8
 Epoxy putty 0.1 ml 13/13
 Heatshrink, black irradiated polyolefin tubing; d: 9.5 mm 2/2
Leads
 Copper wire 1 annealed copper stranded wire, silicone rubber insulation; 128/0.05 2/2
 Copper wire 2 silver plated copper stranded wire, PTFE insulation; 7/0.12 2/2
 Carbon wire 0/0
Electrode adhesives/gels
 Collodion Adhesive: SLE UK 0.1 ml 4/4
 Electrode gel: Dracard ECG Gel, Crown Graphic UK 0.1 ml 9/9
 Elefix electrode paste: Nihon Kohden combined adhesive/gel 0.1 ml 8/6

Ag: silver, AgCl: silverchloride, Au: gold. Dimensions of components: length/width/height, d: diameter. The two values in the third column represent the results of the two measurements.