Inhalation Technique Errors with Metered-Dose Inhalers Among Patients with Obstructive Lung Diseases: A Systematic Review and Meta-Analysis of U.S. Studies

Soojin Cho-Reyes 1, Bartolome R Celli 2, Carole Dembek 3, Karen Yeh 1, Maryam Navaie 1,4
PMCID: PMC6872219  PMID: 31342732

Abstract

Background: Metered dose inhalers (MDIs) are commonly prescribed for inhalation therapy, but correct use is critical to promoting effective medication delivery. This systematic literature review and meta-analysis evaluates the overall and step-by-step prevalence of errors among adults with obstructive lung diseases in the United States who used MDIs.

Methods: Electronic and manual searches conducted between 1979-2018 using PubMed, EMBASE, PsycINFO, Cochrane, and Google identified 10 articles that met the following inclusion criteria: (a) English language, (b) U.S. adults diagnosed with chronic obstructive pulmonary disease, and (c) MDI use error rates. Meta-analytic techniques using random-effects models were applied to calculate effect sizes, weighted proportions, and 95% confidence intervals (CIs). Heterogeneity was assessed by the I2 statistic.

Results: Aggregate findings revealed that 86.7% of patients (n=390, 95% CI 77.5-96.0) made at least 1 inhalation technique error, and 76.9% (n=885, 95% CI 65.8-87.9) incorrectly performed ≥ 20% of device use steps. The most prevalent step-by-step errors across the studies (n=1105) were failure to: (a) exhale fully and away from the inhaler before inhalation (65.5% [95% CI 52.0, 78.9]); (b) hold breath for 5-10 seconds (41.9% [95% CI 29.8, 53.9]); (c) inhale slowly and deeply (39.4% [95% CI 26.2, 52.5]); (d) exhale after inhalation (35.9% [95% CI 17.0, 54.8]); and (e) shake the inhaler before use (34.2% [95% CI 30.6, 37.7]).

Conclusions: Across the studies used in this meta-analysis more than three-fourths of U.S. adults with obstructive lung diseases used MDIs incorrectly. Our findings suggest the need for ongoing patient education and consideration of alternative devices to mitigate errors.

Keywords: copd, chronic obstructive pulmonary disease, asthma, systematic literature review, meta-analysis, inhaler errors, inhalation technique, handheld inhalers, metered-dose inhaler errors, MDI errors, nebulizers, inhaler education

Introduction

Note: An earlier version of this manuscript was presented, in part, at the annual conference of the American College of Chest Physicians, October 6-10, 2018 in San Antonio, Texas.

In the United States, the most prevalent obstructive lung diseases are asthma and chronic obstructive pulmonary disease (COPD).1,2 Taken together, these conditions are a leading cause of morbidity and mortality among U.S. adults.2,3 An estimated 16.5 million Americans are diagnosed with asthma, 16 million with COPD, and 0.94 million with asthma–COPD overlap syndrome.1-4 The economic burden of asthma and COPD is substantial. The costs of treating asthma range between $56 billion5 and $81.9 billion,6 of which 61% is associated with medical costs.6 The health care costs of COPD are similarly staggering, accounting for more than $72.7 billion yearly.7

Although neither asthma nor COPD can be cured, both can be managed through inhaled bronchodilators that deliver aerosolized medications into the lungs. A range of inhaler devices can be used to administer bronchodilators including metered-dose inhalers (MDIs), dry powder inhalers, soft mist inhalers, and nebulizers.8 MDIs are the most commonly used handheld devices,9 but they require patients to have good hand-breath coordination so they can simultaneously manage breathing and actuation for effective drug delivery.10 As a result, MDIs are considered to be inherently more difficult to administer than alternative devices.11 The importance of proper inhaler technique cannot be over emphasized. Past research has shown that incorrect inhaler technique may compromise medication delivery,12-14 resulting in poor outcomes over time including increased risk of exacerbations, higher health resource utilization, and mortality.13-17

Difficulty using MDIs has long been documented in the literature. In 2001, a meta-analysis (MA) of 24 studies found that nearly 77% of international (non-U.S.) patients with asthma or COPD made at least 1 error while using MDIs.18 More recently, another internationally-focused MA study reported that 86.8% of patients with asthma or COPD made errors using an MDI, 45.6% of which were considered to be critical (errors that would directly impact the effectiveness of the drug).11 In an attempt to facilitate MDI use, a spacer is often recommended.8,19 However, to date, there is little evidence that shows the use of a spacer actually reduces inhalation technique errors. Indeed, more than 60% of patients, on average, have been reported to continue having difficulty using MDIs correctly even with a spacer.20-24

The extent and type of inhalation technique errors with MDIs have been synthesized among patients in diverse international settings through previous narrative reviews,25-30 systematic literature reviews (SLRs),12,31-37 and MA studies.11,17,18 However, none of these past studies conducted a step-by-step MA of errors to identify the device use steps patients found to be the most challenging. Moreover, the study populations in previous MA studies have been predominantly globally focused, with limited attention on U.S. patients. Individual studies on U.S. populations with obstructive lung diseases have shown that the proportion of patients making at least 1 inhalation technique error or errors in ≥ 20% of MDI device use steps ranges between 65% to 100%.21-24,39-41 However, drawing inferences from these studies is limited due to the relatively small sample sizes and heterogeneity across populations, error definitions, care settings, and diagnoses. To date, there has been no comprehensive study combining an SLR and MA on MDI errors in U.S. patients to evaluate the prevalence and types of errors.

Differences between U.S. and international patients with obstructive lung diseases have been documented with regards to sociodemographic characteristics (e.g., education and/or literacy levels),42-44 comorbidities and pulmonary disease severity.19,29,42,45 In addition, differences in health system factors (e.g., payer source and physician training) have been reported to play a role in disease management among U.S. and non-U.S. patients with lung disease.19,26,29,45 These differential attributes can influence the patient’s ability to properly use MDIs.35,42,46 Therefore, synthesized findings based on international populations may not be directly generalizable to U.S. patients. The purpose of this study was 2-fold: (a) to quantify the prevalence of inhalation technique errors among U.S. adults with COPD who use MDIs, and (b) to determine which device use steps were the most problematic for patients using MDIs.

Methods

Search Strategy

An SLR was conducted using PubMed, EMBASE, PsycINFO, Cochrane, and Google Scholar. We identified studies published from January 1, 2010 to November 15, 2018 that met the following criteria: (a) English language; (b) U.S. adults diagnosed with COPD; and (c) information available on MDI errors. In a few articles, authors combined patients diagnosed with COPD, asthma, or other chronic obstructive disease into a single group and presented their results on this combined study population. Thus, while we restricted our search criteria to COPD patients, if a publication combined COPD with other diagnoses of obstructive lung disease, that study would have been identified by our search and evaluated further.

The following data were extracted from the studies: (a) patients’ sociodemographic characteristics (e.g., age, sex, health literacy level); (b) primary diagnosis; and (c) inhalation technique and step-by-step device use errors. In addition to electronic searches, a manual search was conducted covering the period of January 1, 1979 to November 15, 2018 by reviewing all of the references contained within the articles identified from electronic searches.

Quality Assessment

Study quality was graded by 3 trained independent reviewers. Studies that met the following criteria were classified as poor quality and excluded from the MA: (a) combined error rates across multiple devices; (b) reported error rates for ≤ 5 patients; and (c) if there was no way to ascertain the percentage of patients who made inhalation technique errors from the data provided. For randomized controlled trials, quality was assessed using criteria published by the Cochrane Collaborative.47 For cross-sectional and observational studies, quality was assessed using applicable scales by Newcastle and Ottawa.48

Metered-Dose Inhaler Device Use Steps

Reported errors were based on the patients’ ability to correctly complete the following 12 device use steps: (1) removing the cap; (2) shaking the inhaler before use; (3) attaching the inhaler to a spacer, as needed; (4) holding the inhaler upright; (5) exhaling completely (and away from the inhaler) before inhalation; (6) placing the inhaler mouthpiece between teeth and sealing lips; (7) actuating once during inhalation; (8) inhaling slowly and deeply; (9) holding breath for 5-10 seconds; (10) removing the inhaler or spacer from the mouth; (11) exhaling and breathing normally; and (12) repeating steps for a second puff. These steps were the most frequently reported across the 10 studies evaluated in this MA. Additional steps that were reported in only 1 study (e.g., checking expiration date and priming device)49 were not analyzed. For the purposes of step-by-step device use error analysis, all steps were included with the exception of 1 and 4 due to limited data availability.

Statistical Analysis

The proportion of MDI device use errors was analyzed using the following 2 definitions depending on data availability: (1) the percentage of patients who made at least 1 inhalation technique error24,41,46,49,50; or (2) the percentage of patients who incorrectly performed ≥ 20% of device use steps.21-24,40 For one study,39 the percentage of patients with at least 1 inhalation technique error was inferred based on the reported device use step with the highest frequency of error. For studies that reported inhalation technique errors for both pre- and post-interventions, only pre-intervention (i.e., baseline) error data were used for both control and intervention groups.

Step-by-step device use errors were also analyzed. For studies that presented error data graphically, numeric values were estimated by interpreting graphic representations and validated by 2 independent analysts. Additional calculations were performed as follows, depending on selected scenarios: (1) if authors presented the percentage of patients who performed each step correctly, results were converted to the percentage of patients who performed each step incorrectly; (2) if authors reported 2 error rates for each device use step, a mean value was computed; and (3) if authors reported error rates for 2 different subgroups instead of their entire study population, weight means were calculated based on the proportion of each subgroup.

Meta-Analysis

Information about the prevalence of overall inhalation technique errors as well as step-by-step device use errors was gathered according to the recommendations outlined by the “Meta-analysis of Observational Studies in Epidemiology” guidelines51 for MA and systematic reviews. Pooled estimates (weighted proportions) and 95% confidence intervals (CIs) were computed using the approximation of a binomial distribution.52 To visually inspect the degree to which the effect estimates of each study distributed around the pooled effect estimates, forest plots were created for pooled and step-by-step device use errors. Heterogeneity was assessed by the I2 statistic to determine the suitability of combining estimates across studies.53,54 I2 values of 25%, 50%, and 75%, corresponded to small, moderate, and large degree of heterogeneity.55 Given the small number of studies, we performed a restricted maximum likelihood random-effects model using JASP 9.2 software.54,55

Results

Search Results

A total of 615 potential publications met the initial screening criteria (Figure 1). The manual search identified an additional 22 publications. After removing duplicate studies (n=193), 444 abstracts were evaluated further to determine relevance. After eliminating 349 abstracts that did not meet the eligibility criteria, 95 full-text articles remained and were reviewed in detail. Among these articles, 85 studies did not have data on U.S. adults or MDI device errors, resulting in a total of 10 publications that were included in this MA.

graphic file with name JCOPDF-6-267-f001.jpg

Patient Characteristics

There were 1360 patients included across the 10 studies (Table 1). The range in the average age of the patients was from 38 to 82 years. More than half of the patients were female (54%) and nearly three-quarters (74%) had a COPD diagnosis. Between 9% to 37% of patients had low health literacy (based on reports from 7 studies). Due to use of multiple devices and participants who did not complete the entire study cycle, MDI errors were reported for a total of 1275 patients, 27.6% (n=352) of whom also used a spacer.

graphic file with name JCOPDF-6-267-t001.jpg

Inhalation Technique Errors Using Metered-Dose Inhalers

Aggregate findings across 6 of 10 studies (n=390) revealed that 86.7% of patients (95% CI 77.5-96.0) made at least 1 inhalation technique error (Figure 2a). Across the remaining 4 of 10 studies (n=885), 76.8% of patients (95% CI 65.8-87.9) incorrectly performed ≥ 20% of device use steps (Figure 2b). Further analyses were conducted to evaluate MDI device use errors with and without a spacer (n=352 across 4 unique studies, and n=923 across 6 unique studies, respectively). A slightly higher proportion of patients who used a spacer made inhalation technique errors as compared to those who did not use a spacer (82.7% [95% CI 73.2-92.2] versus 78.2% [95% CI 61.4-94.9], respectively) (Figures 3a and 3b). I2 values were small (< 50%), except for patients who did not use a spacer (Figure 3b; I2=78.3%).

graphic file with name JCOPDF-6-267-f002a.jpg

graphic file with name JCOPDF-6-267-f002b.jpg

graphic file with name JCOPDF-6-267-f003a.jpg

graphic file with name JCOPDF-6-267-f003b.jpg

Step-By-Step Metered-Dose Inhaler Device Use Errors

Eight of the 10 (80%) studies reported error data for each device use step (n=1105 patients). The most frequent errors were: (a) failure to attach the inhaler to the spacer when required (based on 2 studies using MDI with a spacer, n=24) (78.1% [95% CI 65.3, 91.0]); (b) failure to exhale fully (and away from the inhaler) before inhalation (65.5% [95% CI 52.0, 78.9]); (c) failing to hold breath for 5-10 seconds (41.9% [95% CI 29.8, 53.9]); (d) inhaling too fast and not deeply (39.4% [95% CI 26.6, 52.5]); (e) failing to exhale after inhalation (35.9% [95% CI 17.0, 54.8]); and (f) failing to shake the inhaler before use (34.2% [95% CI 30.6, 37.7]) (Figure 4). Additionally, 43.2% (95% CI 31.3, 55.0) of patients failed to repeat the steps for a second puff. I2 values were negligible for steps 2, 3, and 6. For steps 5 and 7 through 12, I2 ranged from 77.9% to 98.6%.

graphic file with name JCOPDF-6-267-f004.jpg

Discussion

To the best of our knowledge, this is the first SLR and MA to quantify the prevalence of MDI inhalation technique errors and identify the most problematic device use steps for U.S. adults with obstructive lung diseases. Aggregate findings across the identified studies revealed that 8 of 10 patients who used MDIs made at least 1 inhalation technique error. In addition, 7 of 10 patients incorrectly performed at least 20% of MDI device use steps.

Our results are comparable to previously reported findings on international patients. In an SLR and MA study conducted by Chrystyn et al,11 the authors found that 86.8% of patients using MDIs made at least 1 inhalation technique error, of which 45.6% were considered a critical error (defined as an error that may directly impact the effectiveness of the delivered drug). In contrast to both our findings and those reported by Chrystyn et al, Sanchis et al34 reported an average inhalation error rate of about 40% among MDI users in their MA. This lower error rate likely stemmed from Sanchis et al’s more restricted definition of errors which was limited to only device use errors related to coordination and breath-holding inhalation steps. In addition, unlike our study, Sanchis et al’s study population was not limited to adults (e.g., included children with asthma).

Our results show that the most common errors were associated with coordination and breath-holding steps. We found that 2 of 3 patients had difficulty exhaling fully and away from their MDIs before inhalation. Errors that result in not emptying the lungs completely could have negative consequences on appropriately inhaling prescribed medications. We also found about 2 of 5 patients failed to hold their breath for the few seconds needed in order to ensure that the medication was successfully delivered to the lungs. This type of error can negatively impact treatment due to poor medication deposition. Moreover, 1 in 3 patients failed to inhale slowly and deeply. This type of error also adversely affects medication deposition. Other steps involved in dose preparation, such as failing to shake the inhaler before use, led to additional errors in 1 of 3 MDI users. Collectively, our findings seem to indicate lower MDI error rates associated with coordination and breath-holding steps among U.S. patients when compared to international patients, of whom > 45% have been reported to make these types of errors in previous MA studies.11,34

We also found that a higher proportion of patients who used MDIs with a spacer made inhalation technique errors, as compared to patients who did not use a spacer. This finding was surprising and not consistent with the general reason for prescribing a spacer; that is, to facilitate inhalation technique steps that could lead to lower errors in MDI use.8,13,57 Our findings are also not consistent with internationally-focused MA studies which have reported substantially lower MDI device use errors when a spacer is used.11,34 It is not clear why our findings differ. One plausible explanation is that selection bias is playing a role since U.S. clinicians tend to prescribe a spacer for an MDI for patients who are more likely to have difficulty performing inhalation steps correctly compared to patients who demonstrate greater competency using MDIs.13,58,59 Another possibility is related to the fact that all of the patients who used an MDI with a spacer came from multiple studies conducted by Press et al. It is plausible that those patients could have been from the same institution and potentially not independent samples. Yet a third plausible explanation could be the differential age distribution in the study samples, a critical factor shown to be associated with MDI device use errors and the probability of using a spacer.35,36

The predominance of errors documented across the many studies included in this SLR and MA underscores the importance of patient training and education on the proper use of MDIs. Indeed, global respiratory guidelines for both asthma and COPD suggest that participants be instructed on proper use of inhaler devices and also have their inhalation technique re-assessed at each office visit.13,14 Unfortunately, past research has shown that between 25% to 30% of patients never receive instructions on inhaler use.50,60 Moreover, up to 69% of patients say that their health care provider has never observed them using their inhalers.61 Studies have also shown that only 45% of health care providers assess device technique in every newly diagnosed patient.62 Taken together, it is not surprising that many patients report not feeling confident that they are using their inhalers correctly.63-65 Even when patients feel confident that their inhalation technique is correct, there is value in health care providers continually reassessing the patients’ technique. Indeed, multiple studies have shown that the probability of patients correctly performing inhalation steps increases with the higher number of demonstrations by health care providers.42,64,66 Clearly, educating patients on proper inhaler technique is of paramount importance in reducing the risk of inhalation errors.13,17,66

Although a recent SLR of educational interventions found that > 90% of studies reported significant improvements in the patients’ inhalation technique after they received training, education on device use is not an adequate enough strategy to reduce technique errors for some patients.37 For MDIs, past studies have reported persistently high rates of incorrect MDI use even after training and educational interventions.32,67,68 In real-world settings, incorrect MDI use remains challenging because compared to other types of devices, MDIs require more hand-breath coordination, higher cognitive abilities to operate, the ability to hold breath, and the need to generate adequate inspiratory force.29 These types of requirements have led to some patients ceasing inhaled therapy due to the perceived complexity of using devices.69 Thus, for some patients, inhalation technique errors may be mitigated by using alternative devices, such as nebulized delivery that does not require hand-mouth coordination or breath holding for proper medication administration. Past studies have shown that many patients with difficulty using handheld inhalers have positive perceptions of nebulizer use.64 Ultimately, device features such as simplicity and ease of use, convenience, and overall experience are important to patients.69,70 Therefore, sensitizing health care providers to more carefully select inhalation devices that suit patients’ abilities and preferences, while meeting their medical needs, is imperative to reducing device use errors.

Our results should be considered in light of certain study limitations. First, a relatively small number of studies met our inclusion criteria. The number of studies with data available on step-by-step device use errors was even smaller. Thus, our findings may have limited generalizability to larger populations of asthma or COPD patients. Second, there was little to no information available on the association between MDI device use errors and the clinical consequences of errors on symptom management and health outcomes. Third, we were unable to examine a multitude of factors that have been shown to influence the incidence of MDI device use errors, such variability in the level of training patients may have received from providers, the patient’s physical or cognitive limitations, the subjectivity of the assessor’s technique in evaluating patient errors, and patient-introduced bias (i.e., the Hawthorne effect).71 Fourth, we were unable to distinguish potential differences in error rates between asthma and COPD patients because the majority of the studies combined the 2 diagnoses when reporting MDI use errors. Lastly, our results on inhalation technique errors among patients who used MDIs with a spacer may have been influenced by selection bias since all of the studies were published by the same group of authors in 1 setting; therefore, the uniqueness of the study samples could not be ascertained.

Conclusions

In this SLR and MA, we found that more than 8 out of 10 patients in our sample of U.S. adults with obstructive lung diseases made at least 1 inhalation technique error using MDIs. At least 1 out of 3 patients had difficulty exhaling fully (and away) before inhalation, holding their breath, and inhaling slowly and deeply. Our findings suggest the need for ongoing patient education and consideration of alternative devices, such as nebulizers, that may mitigate technique errors by eliminating some of the inhalation steps that patients have difficulty completing.