Figure 4.

Figure 4

Bland-Altman plot of the differences between tidal volume measured by the turbine driven ventilator and the lung simulator in the obstructive condition. Data are reported in ml. PEEP: Positive end expiratory pressure (values are expressed in cmH2O).TDV: tidal volume (VT) measured by turbine driven ventilator; LS: Tidal volume (VT) measured by lung simulator. Average (x-axys): average of the two measurements; TDV – LS (y-axys): difference between mesurements. Bias: Average of the differences between measurements. 95% LoA: Limits of agreement (±1.96 standard deviation - SD - of the bias). In this plot, the overall bias is 141 ml, and 95% LoA are from 109 to 173 ml. Although all plotted differences are within the 95% LoA, a large clinical relevance bias is found at all tested PEEP levels.