Differences in tidal volumes (VT) measured by turbine driven ventilator and lung simulator at different levels of PEEP in the bench study.

Simulated Condition (TDV-LS) PEEP 5 cmH2O (TDV-LS) PEEP 8 cmH2O (TDV-LS) PEEP 10 cmH2O (TDV-LS) PEEP 12 cmH2O P value
Normal 51 (3)°+ 31.1(3.5)*+ 21.8 (2)*° <0.001
Restrictive 61 (3)° 10,4 (1.3)* 1,1 (1.6)*°§ −11,9 (1.9)*°+ <0.001
Obstructive 164 (2)°+ 132,5 (1.6)*+ 127,3(3.1)*° <0.001

Data are expressed in ml and reported as mean (±SD).

PEEP: positive end expiratory pressure; TDV: turbine driven ventilator; LS: lung simulator; (TDV-LS): difference between VT measurements by turbine driven ventilator and lung simulator.

*Different from 5; °different from 8, +different from 10, §different from 12.