Table 1.
Evaluation of MR thermometry quality ¡n patients with solid tumors in various locations.
Location/Type | Age/Sex | Sedation | Accuracy [°C] |Tmean-371 | Uniformity [°C] Tsd, spatial | Stability [°C] Tsd, temporal |
---|---|---|---|---|---|
Femur ES | 11 F | None | 0.12 | 0.34 | 0.28 |
Tibia OS | 14 M | None | 0.44 | 1.55 | 1.10 |
Tibia OS | 10 M | None | 0.18 | 0.93 | 0.76 |
Pelvis NB | 4 F | Sedateci | 0.28 | 0.53 | 0.54 |
Pelvis ES | 11 F | None | 3.89 | 4.00 | 7.03 |
Rib ES | 12 M | Sedated | 1.58 | 2.95 | 2.53 |
Spine MPNST | 10 M | Sedated | 1.86 | 3.92 | 3.56 |