Table 3.

Comparison of the accuracy and stability of MR thermometry evaluated in this study with other references (A) and the estimated diffusivity generated with our model compared with other studies (B).

A Study Model Accuracy (°C) Stability (°C)
Our Study Pigs 0.6 ± 0.2 0.7 ± 0.2
Phantom 0.4 ± 0.1 0.5 ± 0.1
Tillander Pigs −0.7 ± 0.9 0.8
Chu Pigs −0.2 ± 0.3 0.6 ± 0.5
Volunteer 1.1 ± 0.4 0.7 ± 0.4
Zhang Pigs N/A 1.1 ± 0.2
Patients N/A 1.2 ± 0.4
B Study Model Diffusivity (mm2/s)
Our Study Pigs 0.17
Phantom 0.16
Dragonu Porcine kidney 0.17 ± 0.00
Dillon Porcine thigh 0.15 ± 0.02
0.15 ± 0.01
Rabbit 0.15 ± 0.01
Phantom 0.15 ± 0.00
Zhang Pigs 0.14 ± 0.01
Zhang Patients 0.12 ± 0.02