. 2018 Nov 30;5(2):28–37. doi: 10.14338/IJPT-17-0035.1

Table 4.

The overall uncertainty of the calibration curves. There were 6 points of measurements, which are shown in the “Relative location” column. For the averaged computed tomography (CT) calibration evaluation, the averaged CT calibration curve was applied for all 5 CT images to obtain the water-equivalent thickness (WET) values from the treatment planning system (TPS). For the CT-specific calibration evaluation, the CT-specific calibration curves were applied to a given CT scan to obtain the WET values from the TPS.

Relative location (cm)
Averaged CT calibration (%)
CT-specific calibration (%)
CT1
CT2
CT3
CT4
CT5
CT1
CT2
CT3
CT4
CT5
X = 0 −2.8 −0.6 −2.3 −0.1 −1.2 −3.0 −0.5 −2.5 −1.3 −1.6
X = −4.1 −3.7 −3.0 −3.7 −3.3 −3.5 −4.3 −4.0 −3.3 −3.4 −3.7
X = −8.4 −2.2 −0.7 −0.8 1.2 0.1 −2.2 −0.8 −1.7 −0.1 −0.2
X = −13.4 −2.1 1.2 0.5 1.6 1.2 −1.5 0.9 0.6 1.8 0.9
X = −17.9 −0.2 0.1 −0.4 0.3 −0.6 −1.9 −0.7 −0.6 −0.7 −0.8
Z = 2 2.4 2.6 2.6 3.4 2.8 1.9 2.0 2.3 2.4 2.3
RMSE 2.5 1.7 2.1 2.1 2.0 2.7 1.9 2.1 2.0 2.0

Abbreviation: RMSE, root mean square error.