(a) Output voltage (after removing the gain of 80 dB) recorded from the Chebyshev
notch channel when a sinusoidal single tone (25 Hz, amplitude 100 nV peak) was
injected to the input of the channel. (b) Output voltage recorded from the
Powerlab 26T bioamplifier when a sinusoidal single tone (25 Hz, amplitude 100 nV
peak) was injected to the input of the system. (c) Amplitude spectrum calculated
when two sinusoidal tones, one low-frequency (=5 Hz) and one higher-frequency
(=25 Hz) are sequentially injected to the inputs of the two AFEs. The amplitude
spectrums of both systems present two spectral peaks at 5 and 25 Hz, which are
characterized by the same amplitude. (d) Output voltage (after removing the gain
of 80 dB) recorded from the Chebyshev notch channel when a sinusoidal single
tone (25 Hz, amplitude 30 nV peak) was injected to the input of the channel. (e)
Output voltage recorded from the Powerlab 26T bioamplifier when a sinusoidal
single tone (25 Hz, amplitude 30 nV peak) was injected to the input of the
system. (f) Amplitude spectrum calculated when two sinusoidal tones, one
low-frequency (=5 Hz) and one higher-frequency (=25 Hz), are sequentially
injected to the inputs of the two AFEs. The amplitude spectrums of both systems
present two spectral peaks at 5 and 25 Hz, which are characterized by the same
amplitude.